12 Lead Digital Electrocardiograph Using an ADS1298 Converter and Digital Filters

Authors

  • Ariel Quezada

DOI:

https://doi.org/10.52428/20758944.v13i40.648

Keywords:

Electrocardiograph, 12 lead, Digital signa! processing

Abstract

The development of a 12-lead electrocardiograph using analog filters involves the use of a large number of components and presents problems related to the variation of their nominal values, their tolerances and effects of time and temperature. The proposed system uses an ADS1298 24-bit resolution converter, a microcontroller a n d a personal computer, used to capture the information and save it in an open format. This information is processed using digital filters to remove noise signa Is at specific frequencies. When using digital filters, a smaller amount of components is used and the effect of their tolerances and degradation in the behavior of the filters is avoided. The system has been developed using only free and open source programming tools.

Downloads

Download data is not yet available.

References

Texas lnstruments. (2010). Low•Power, 8•Channel, 24•Bit Analog Front•End for Biopotential Measurements (Rev. J). 01/2015., Sitio web: http://www.ti.com/lit/ds/symlink/ads1298.pdf

Texas lnstruments. (2010). Analog Front•End Design for ECG Systems Using Delta•Sigma ADCs (Rev. A). 01/2015. Sitio web: http://www.ti.com/lit/an/sbaa160a/sbaa160a.pdf.

Texas lnstruments. (2010}. ADS1298ECG•FE/ADS1198ECG•FE ECG Front•End Performance Demonstration Kit User's Guide. Rev. 01/2016. Sitio web: http://www.ti.com/lit/ug/sbaul 71c/sbaul 71c.pdf.

Unpingco, J. (2014). Python for Signa[ Processing. Suiza: Springer. https://doi.org/10.1007/978-3-319-01342-8

Torrez, M. y Quezada, A. (2016}.Diseñode un Sistema Medidor de Vibraciones para Máquinas Rotativas de 900 rpm hasta 84000 rpm. Journal Boliviano de Ciencias. 12, 23•28. https://doi.org/10.52428/20758944.v12i37.671

Camacho, A., Quezada, A. y Morales, l. (2016). Sistema de Adquisición y Tratamiento Digital de las Doce Derivaciones Cardiacas del Electrocardiograma. Journal Boliviano de Ciencias. 12, 29•33. https://doi.org/10.52428/20758944.v12i37.672

Published

30-08-2017

How to Cite

Quezada, A. (2017). 12 Lead Digital Electrocardiograph Using an ADS1298 Converter and Digital Filters. Journal Boliviano De Ciencias, 13(40), 48–53. https://doi.org/10.52428/20758944.v13i40.648

Issue

Section

Scientific Paper